Test Parameter |
|Z|,|Y|,C,L,X,R,B,G,D,Q,θ,DCR,Turn-Ratio,Phase |
|
Test Frequency |
Range:
20 Hz ~300 kHz;Resolution: 10mHz;Accuracy: 0.01% |
|
Test Level |
Range:
5 mVrms~2 Vrms, Resolution:1mV; 50 µArms ~20 mArms, Resolution:10µA;
Accuracy: 5% |
|
Θ |
Degree:
-179.999ş to 179.999ş
Radian: -3.14159 to 3.14159 |
|
Measurement Speed |
Fast:32ms,
Medium:90ms, Slow:650ms |
|
Trigger |
Internal, Manual, External, Bus(HP-IB or RS232C) |
|
Correction Function |
OPEN,
SHORT and LOAD selectable |
|
Internal DC Bias Source |
0 V,
1.5 V, 2 V;Accuracy: 1% |
|
Comparator |
The
built-in comparator can sort devices into a maximum of ten bins (BIN 1
to BIN 9 and the OUT OF BINS bin) using a maximum of nine pairs of
primary limits and one pair of secondary parameter limits. |
|
Input Power |
115V /
230Vac±10%, 50Hz / 60Hz, Fuse 4A / 2A |
|
Interface |
Standard RS232 and USB interface, Optional GPIB (IEEE488),HANDLER and
SCANNER interface |
|
1801 4 terminals test fixture |
x1 |
|
1802 4 terminal Kelvin test clip leads
|
x1 |
|
1803 2 terminal test clip leads |
x1 |
|
1804 Glided shorting plate |
x1 |