8330 - Analisador Automático de Componentes

Model
8330
Test Parameter |Z|,|Y|,C,L,X,R,B,G,D,Q,θ,DCR,Turn-Ratio,Phase
Test Frequency Range: 20 Hz ~300 kHz;Resolution: 10mHz;Accuracy: 0.01%
Test Level Range: 5 mVrms~2 Vrms, Resolution:1mV; 50 µArms ~20 mArms, Resolution:10µA; Accuracy: 5%
Output Impedance 30Ω,100Ω
Basic Accuracy 0.05%
Display Range
|Z|,R,X
0.00001Ω~99.9999MΩ
DCR 1mΩ~99.9999Ω
|Y|,G,B 0.00001μS~99.9999S
C 0.00001pF~9.99999F
L 0.00001μH~9999.99H
D 0.00001~9.99999
Q 0.00001~99999.9
Θ Degree: -179.999ş to 179.999ş
Radian: -3.14159 to 3.14159
Turns-Ratio 0.001 to 99.9999M
Δ% -999.999%~999.999%
Measurement Speed Fast:32ms, Medium:90ms, Slow:650ms
Equivalent Circuit Mode Series, Parallel
Ranging Mode Auto / Manual (Hold)
Trigger Internal, Manual, External, Bus(HP-IB or RS232C)
Averaging Rate 1 ~ 255, programmable
Correction Function OPEN, SHORT and LOAD selectable
Internal DC Bias Source 0 V, 1.5 V, 2 V;Accuracy: 1%
Comparator The built-in comparator can sort devices into a maximum of ten bins (BIN 1 to BIN 9 and the OUT OF BINS bin) using a maximum of nine pairs of primary limits and one pair of secondary parameter limits.
General
Input Power 115V / 230Vac±10%, 50Hz / 60Hz, Fuse 4A / 2A
Memory 20 Memories
Display 320 X 240 Graphic LCD
Interface Standard RS232 and USB interface, Optional GPIB (IEEE488),HANDLER and SCANNER interface
Environment 0~40℃, ≤90%RH
Dimension 430 X 186 X 490
Net Weight 15Kg
Standard Accessories
Power Cord x1
1801 4 terminals test fixture x1
1802 4 terminal Kelvin test clip leads x1
1803 2 terminal test clip leads x1
1804 Glided shorting plate x1